The impact of pitch values on image quality and radiation dose in an abdominal adult phantom using CT

dc.contributor.authorLança, Luís
dc.contributor.authorBarros, Pietro
dc.contributor.authorD'Agostini Derech, Rodrigo
dc.contributor.authorHiggins, Daniel
dc.contributor.authorKleiker, Marjolein
dc.contributor.authorLiardet, Sébastien
dc.contributor.authorLøvlien, Ine Michaela
dc.contributor.authorMcNally, Kevin
dc.contributor.authorThévenaz, Manon
dc.contributor.authorHogg, Peter
dc.date.accessioned2017-08-16T11:48:56Z
dc.date.available2017-08-16T11:48:56Z
dc.date.issued2016
dc.description.abstractPurpose: To identify the impact of different pitch values on image quality and effective radiation dose for axial and coronal plane in abdominal adult CT. Methods and materials: Three scans were conducted on an abdominal phantom using a Toshiba Aquilion 16-slice CT scanner with three different pitch values: standard (0.938), detail (0.688) and fast (1.438). Slices were taken from the upper, middle and lower abdomen in the axial plane and anterior, middle and posterior in the coronal plane. The six different anatomical structures were liver, intrahepatic vessels, spleen, pancreas, kidneys and renal vessels, retroperitoneum, aorta and vena cava. A two-alternative forced-choice (2AFC) method was used to evaluate images for each pitch with 8 observers using a 3-point Likert scale. SNR was calculated in every plane, slice and pitch factor using the ImageJ software. To estimate effective radiation dose the CT Expo software was used. Results: Detail pitch factor provides superior image quality compared to standard in axial plane when evaluating the liver (p<0.034) and pancreas (p=0.008). However, the results for spleen, kidney, renal vessels, retroperitoneum, aorta and vena cava are not significant when comparing detail vs standard. Standard provides a 26.3% reduction in effective radiation dose (mSv) compared to detail. Fast had the worst image quality in both the axial and coronal plane but the lowest dose. In the coronal plane, standard was superior to both detail (p=0.026) and fast (p=0.023) in terms of image quality. The differences in SNR results were not significant except in standard vs detail in the coronal plane (p=0.03). Conclusion: Detail pitch factor provides superior image quality to standard and fast in the axial plane. Standard had superior image quality to both detail and fast in the coronal plane. The augmentation of effective doses has been inversely proportional to the pitch factors. The most irradiant pitch mode was detail and the less was fast.pt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.citationLança L, Barros P, D’Agostini Derech R, Higgins D, Kleiker M, Liardet S, et al. The impact of pitch values on image quality and radiation dose in an abdominal adult phantom using CT. In: OPTIMAX 2016 – Optimising image quality for medical imaging: University of Salford, Salford, Greater Manchester, UK. Manchester: University of Salford; 2016. p. 105-16.pt_PT
dc.identifier.isbn9781907842931
dc.identifier.urihttp://hdl.handle.net/10400.21/7322
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.publisherUniversity of Salfordpt_PT
dc.relation.publisherversionhttp://usir.salford.ac.uk/41428/pt_PT
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/pt_PT
dc.subjectComputed tomographpt_PT
dc.subjectAbdomenpt_PT
dc.subjectPitch factorpt_PT
dc.subjectEffective radiation dosept_PT
dc.subjectImage qualitypt_PT
dc.subjectPhantom studypt_PT
dc.titleThe impact of pitch values on image quality and radiation dose in an abdominal adult phantom using CTpt_PT
dc.typebook part
dspace.entity.typePublication
oaire.citation.endPage116pt_PT
oaire.citation.startPage105pt_PT
oaire.citation.titleOPTIMAX 2016 – Optimising image quality for medical imagingpt_PT
rcaap.rightsopenAccesspt_PT
rcaap.typebookPartpt_PT

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
The impact of pitch values on image quality and radiation dose in an abdominal adult phantom using CT.pdf
Size:
12.6 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: