Repository logo
 
Publication

Failure on clinical image quality criteria in digital mammography: how can radiographers do better?

dc.contributor.authorReis, Cláudia
dc.contributor.authorOliveira, Mário
dc.contributor.authorPascoal, Ana
dc.contributor.authorAlves, João
dc.date.accessioned2015-08-17T15:17:51Z
dc.date.available2015-08-17T15:17:51Z
dc.date.issued2014-03
dc.description.abstractPurpose: To assess image quality using PGMI (perfect, good, moderate, inadequate) scale in digital mammography examinations acquired in DR systems. Identify the main failures and propose corrective actions. Evaluate the most typical breast density. Methods and Materials: Clinical image quality criteria were evaluated considering mammograms acquired in 13 DR systems and classified according to PGMI scale using the criteria described in European Commission guidelines for radiographers. The breast density was assessed according to ACR recommendations. The data were collected on the acquisition system monitor to reproduce the daily practice of the radiographer. Results: The image quality criteria were evaluated in 3044 images. The criteria were fully achieved in 41% of the images that were classified as P (perfect), 31 % of the images were classified as M (moderate), 20% G (good) and 9% I (inadequate). The main cause of inadequate image quality was absence of all breast tissue in the image, skin folders in the pectoral muscle and in the infra-mammary angle. The higher number of failures occurred in MLO projections (809 out of 1022). The most represented (36%) breast type was type 2 (25-50% glandular tissue). Conclusion: Incorrect radiographic technique was frequently detected suggesting potential training needs and poor communication between the team members (radiographer and radiologists). Further correlations are necessary to identify the main causes for the failures, namely specific education and training in digital mammography and workload.por
dc.identifier.citationReis C, Oliveira MF, Pascoal A, Pascoal A, Alves J. Failure on clinical image quality criteria in digital mammography: how can radiographers do better? In European Congress of Radiology, Vienna (Austria), 6th to 10th March 2014. Insights Imaging. 2014;5(1 Suppl):S194.por
dc.identifier.doi10.1007/s13244-014-0317-5
dc.identifier.urihttp://hdl.handle.net/10400.21/4781
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherSpringerpor
dc.relation.publisherversionhttp://link.springer.com/article/10.1007/s13244-014-0317-5por
dc.subjectRadiologypor
dc.subjectDigital mammographypor
dc.subjectImage qualitypor
dc.subjectTraining and educationpor
dc.titleFailure on clinical image quality criteria in digital mammography: how can radiographers do better?por
dc.typeconference object
dspace.entity.typePublication
oaire.citation.endPageS194por
oaire.citation.startPageS194por
oaire.citation.titleInsights Imagingpor
oaire.citation.volume5por
rcaap.rightsopenAccesspor
rcaap.typeconferenceObjectpor

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Failure on clinical image quality criteria in digital mammography.pdf
Size:
910.6 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: