Browsing by Author "Pascoal, Ana"
Now showing 1 - 6 of 6
Results Per Page
Sort Options
- Avaliação da dose glandular média (MGD) em três sistemas mamográficosPublication . Reis, Cláudia; Marques, Carla; Lança, Luís; Pascoal, Ana; Pinto, Elsa; Santo, AnaEste estudo tem por objectivos determinar a Dose Glandular Média - Mean Glandular Dose (MGD) - em 3 sistemas de Mamografia e comparar os valores obtidos com os referenciais internacionais. O estudo foi realizado num sistema analógico de Écran-Película (EP) e em dois sistemas de imagem digital (CR e DR). Foi efectuado o cálculo da Entrance Surface Air Kerma (ESAK) e da MGD em três equipamentos a partir de uma amostra de dados referentes a 30 mulheres assintomáticas, com idades compreendidas entre os 40 e 64 anos. Em cada equipamento objecto de análise, foram recolhidos os dados referentes a 10 mulheres. Foram consideradas as projecções crânio-caudal (CC) e oblíqua médio-lateral (MLO). A análise de resultados revelou que o valor de MGD varia quando se compara os três sistemas. Nas incidências CC os valores de MGD obtidos foram de 1,54 mGy (EP), 1,78 mGy (CR) e 0,82 mGy (DR). Nas incidências MLO o valor de MGD foi de 1,53 mGy no sistema EP, de 1,78 mGy no CR e 0,87 mGy no sistema DR. Constata-se que o valor de MGD na incidência de CC é inferior ao valor de MGD na incidência MLO, excepto para o sistema EP. Verifica-se também que o sistema EP apresenta maior variabilidade nos dados de MGD comparativamente com os restantes sistemas. O sistema DR é o que apresenta a menor variabilidade de valores MGD e também valores de MGD mais baixos. Comparando os resultados deste estudo com as referências internacionais, verifica-se que a MGD se encontra abaixo do limite de 2 mGy recomendado. ABSTRACT - This study aims to estimate the Mean Glandular Dose (MGD) associated with three different mammographic systems and compare the results with recommended international reference values. The systems included in the study included a conventional Screen-Film (SF) system and two digital mammography systems (CR and DR). Entrance Surface Air Kerma (ESAK) and MGD associated with each equipment were calculated. A sample of 30 healthy women (age ranging from 40 to 64 years old) undertaking screening mammography was considered in this study. The mammographic exam includes two projections, cranio-caudal (CC) and medio-lateral oblique (MLO). The MGD results obtained for CC projection were 1,54 mGy (SF), 1,78 mGy (CR) and 0,82 mGy (DR). MGD values for the MLO projection were 1,53 mGy (SF), 1,78 mGy (CR) and 0,87 mGy (DR). Results show that MGD value is slightly lower in the CC projection than in MLO, except for the SF system (1,54 mGy; 1,53 mGy). In addition the MGD for the SF system varied more than that associated with the digital systems. The DR system allows a narrow variation of MGD values and also lower MGD values. Comparing this study results with the international references we concluded that MGD values are below the 2 mGy recommended value for the three systems evaluated.
- Caracterização tecnológica da mamografia digital em Portugal: progresso de um projeto em implementaçãoPublication . Reis, Cláudia; Oliveira, Mário; Alves, João; Marques, José Carlos; Pascoal, AnaIntrodução – A mamografia é o principal método de diagnóstico por imagem utilizado no rastreio e diagnóstico do cancro da mama, sendo a modalidade de imagem recomendada em vários países da Europa e Estados Unidos para utilização em programas de rastreio. A implementação da tecnologia digital causou alterações na prática da mamografia, nomeadamente a necessidade de adaptar os programas de controlo de qualidade. Objetivos – Caracterizar a tecnologia instalada para mamografia em Portugal e as práticas adotadas na sua utilização pelos profissionais de saúde envolvidos. Concluir sobre o nível de harmonização das práticas em mamografia em Portugal e a conformidade com as recomendações internacionais. Identificar oportunidades para otimização que permitam assegurar a utilização eficaz e segura da tecnologia. Metodologia – Pesquisa e recolha de dados sobre a tecnologia instalada, fornecidos por fontes governamentais, prestadores de serviços de mamografia e indústria. Construção de três questionários, orientados ao perfil do médico radiologista, técnico de radiologia com atividade em mamografia digital e técnico de radiologia coordenador. Os questionários foram aplicados em 65 prestadores de serviços de mamografia selecionados com base em critérios de localização geográfica, tipo de tecnologia instalada e perfil da instituição. Resultados – Foram identificados 441 sistemas para mamografia em Portugal. A tecnologia mais frequente (62%) e vulgarmente conhecida por radiografia computorizada (computed radiography) é constituída por um detector (image plate) de material fotoestimulável inserido numa cassete de suporte e por um sistema de processamento ótico. A maioria destes sistemas (78%) está instalada em prestadores privados. Aproximadamente 12% dos equipamentos instalados são sistemas para radiografia digital direta (Direct Digital Radiography – DDR). Os critérios para seleção dos parâmetros técnicos de exposição variam, observando-se que em 65% das instituições são adotadas as recomendações dos fabricantes do equipamento. As ferramentas de pós-processamento mais usadas pelos médicos radiologistas são o ajuste do contraste e brilho e magnificação total e/ou localizada da imagem. Quinze instituições (em 19) têm implementado um programa de controlo de qualidade. Conclusões – Portugal apresenta um parque de equipamentos heterogéneo que inclui tecnologia obsoleta e tecnologia “topo de gama”. As recomendações/guidelines (europeias ou americanas) não são adotadas formalmente na maioria das instituições como guia para fundamentação das práticas em mamografia, dominando as recomendações dos fabricantes do equipamento. Foram identificadas, pelos técnicos de radiologia e médicos radiologistas, carências de formação especializada, nomeadamente nas temáticas da intervenção mamária, otimização da dose e controlo da qualidade. A maioria dos inquiridos concorda com a necessidade de certificação da prática da mamografia em Portugal e participaria num programa voluntário. ABSTRACT - Introduction – Mammography is the gold standard for screening and imaging diagnosis of breast disease. It is the imaging modality recommended by screening programs in various countries in Europe and the United States. The implementation of the digital technology promoted changes in mammography practice and triggered the need to adjust quality control programs. Aims –Characterize the technology for mammography installed in Portugal. Assess practice in use in mammography and its harmonization and compliance to international guidelines. Identify optimization needs to promote an effective and efficient use of digital mammography to full potential. Methodology – Literature review was performed. Data was collected from official sources (governmental bodies, mammography healthcare providers and medical imaging industry) regarding the number and specifications of mammography equipment installed in Portugal. Three questionnaires targeted at radiologists, breast radiographers and the chief-radiographer were designed for data collection on the technical and clinical practices in mammography. The questionnaires were delivered in a sample of 65 mammography providers selected according to geographical criteria, type of technology and institution profile. Results – Results revealed 441 mammography systems installed in Portugal. The most frequent (62%) technology type are computerized systems (CR) mostly installed in the private sector (78%). 12% are direct radiography systems (DDR). The criteria for selection of the exposure parameters differ between the institutions with the majority (65%) following the recommendations from the manufacturers. The use of available tools for post-processing is limited being the most frequently reported tools used the contrast/ brightness and Zoom or Pan Magnification tools. Fifteen participant institutions (out of 19) have implemented a quality control programme. Conclusions – The technology for mammography in Portugal is heterogeneous and includes both obsolete and state of the art equipment. International guidelines (European or American) are not formally implemented and the manufacturer recommendations are the most frequently used guidance. Education and training needs were identified amongst the healthcare professionals (radiologists and radiographers) with focus in the areas of mammography intervention, patient dose optimization and quality control. The majority of the participants agree with the certification of mammography in Portugal.
- Failure on clinical image quality criteria in digital mammography: how can radiographers do better?Publication . Reis, Cláudia; Oliveira, Mário; Pascoal, Ana; Alves, JoãoPurpose: To assess image quality using PGMI (perfect, good, moderate, inadequate) scale in digital mammography examinations acquired in DR systems. Identify the main failures and propose corrective actions. Evaluate the most typical breast density. Methods and Materials: Clinical image quality criteria were evaluated considering mammograms acquired in 13 DR systems and classified according to PGMI scale using the criteria described in European Commission guidelines for radiographers. The breast density was assessed according to ACR recommendations. The data were collected on the acquisition system monitor to reproduce the daily practice of the radiographer. Results: The image quality criteria were evaluated in 3044 images. The criteria were fully achieved in 41% of the images that were classified as P (perfect), 31 % of the images were classified as M (moderate), 20% G (good) and 9% I (inadequate). The main cause of inadequate image quality was absence of all breast tissue in the image, skin folders in the pectoral muscle and in the infra-mammary angle. The higher number of failures occurred in MLO projections (809 out of 1022). The most represented (36%) breast type was type 2 (25-50% glandular tissue). Conclusion: Incorrect radiographic technique was frequently detected suggesting potential training needs and poor communication between the team members (radiographer and radiologists). Further correlations are necessary to identify the main causes for the failures, namely specific education and training in digital mammography and workload.
- Overview of the radiographers’ practice in 65 healthcare centers using digital mammography systems in PortugalPublication . Reis, Cláudia; Pascoal, Ana; Radu, Lucian; de Oliveira, Mário Fartaria; Alves, JoãoPurpose: To assess current practices in digital mammography (DM) in Portuguese healthcare providers using digital systems. To investigate compliance with European standards regarding mean glandular dose and quality control practice and to identify optimisation needs. Methods: Two questionnaires, targeted at breast radiographers and chief radiographers, were designed and applied in 65 imaging departments offering DM. Questions fielded were focused on the staff profile and technical/clinical practice. Results: Prior to starting their activity in DM, 70% (82 out of 118) of the respondents received training in DM. The practice in 29 out of 59 providers was established by the manufacturers’ recommendations for image acquisition. Variations were observed between radiographers who belong to the same provider namely the selection of exposure parameters such as the target-filter combination and automatic mode. The use of the manual exposure mode was reported for imaging breast implants (44%) and surgical specimens (22%). The main causes of repeat examinations were skin folding (21%) and absence of pectoral muscle (PM) (20%). Conclusions: The study revealed opportunities to optimise radiographers’ practice in DM regarding the selection of exposure parameters. A robust and consistent training programme in DM and established local protocols can help to reduce the variations observed and improve clinical practice.
- Portuguese study of mean glandular dose in mammography and comparison with European referencesPublication . Reis, Cláudia; Fartaria, Mário João; Alves, João Garcia; Pascoal, AnaTo characterize the mean glandular dose (MGD) in a sample of healthcare providers for digital mammography in Portugal. To compare the achieved values with European references. The MGD was measured on a poly-methyl-methacrylate phantom (45 mm) for each system using dosimeters. In addition, MGD was estimated using exposure settings collected from mammography exams in a clinical context. Data were collected from 25 computed-radiography systems (CR) and 13 integrated digital (DR). For both measurements (phantom and clinical exposures), the average MGD for CR was higher compared to the DR. For CR the mean MGD was 1.85 mGy (CC projection) and 2.10 mGy (MLO projection). For DR systems the corresponding values were 1.54 mGy (CC) and 1.68 mGy (MLO). The average MGD obtained using both methods and for both technologies is within the acceptable reference range proposed by European guidelines (<2.5 mGy). Dose Reference Levels implementation should be the next step to optimise mammography practice in Portugal.
- Quality assurance and quality control in mammography: a review of available guidance worldwidePublication . Reis, Cláudia; Pascoal, Ana; Sakellaris, Taxiarchis; Koutalonis, ManthosObjectives - Review available guidance for quality assurance (QA) in mammography and discuss its contribution to harmonise practices worldwide. Methods - Literature search was performed on different sources to identify guidance documents for QA in mammography available worldwide in international bodies, healthcare providers, professional/scientific associations. The guidance documents identified were reviewed and a selection was compared for type of guidance (clinical/technical), technology and proposed QA methodologies focusing on dose and image quality (IQ) performance assessment. Results - Fourteen protocols (targeted at conventional and digital mammography) were reviewed. All included recommendations for testing acquisition, processing and display systems associated with mammographic equipment. All guidance reviewed highlighted the importance of dose assessment and testing the Automatic Exposure Control (AEC) system. Recommended tests for assessment of IQ showed variations in the proposed methodologies. Recommended testing focused on assessment of low-contrast detection, spatial resolution and noise. QC of image display is recommended following the American Association of Physicists in Medicine guidelines. Conclusions - The existing QA guidance for mammography is derived from key documents (American College of Radiology and European Union guidelines) and proposes similar tests despite the variations in detail and methodologies. Studies reported on QA data should provide detail on experimental technique to allow robust data comparison. Countries aiming to implement a mammography/QA program may select/prioritise the tests depending on available technology and resources.