Image quality and dose analysis for a PA chest X-ray: comparison between AEC mode acquisition and manual mode using the 10 kVp ‘rule’

dc.contributor.authorReis, Cláudia
dc.contributor.authorGonçalves, João
dc.contributor.authorKlompmaker, Corrie
dc.contributor.authorBárbara, Ana Rita
dc.contributor.authorBloor, Chloe
dc.contributor.authorHegarty, Ryan
dc.contributor.authorLagrange, Tania
dc.contributor.authorTemming, Noëlle
dc.contributor.authorSønnesyn, Mathilde
dc.contributor.authorRøkeness, Henriette
dc.contributor.authorYamasathien, Amandine
dc.contributor.authorHogg, Peter
dc.date.accessioned2014-08-07T16:29:50Z
dc.date.available2014-08-07T16:29:50Z
dc.date.issued2014-06
dc.description.abstractPurpose - To compare the image quality and effective dose applying the 10 kVp rule with manual mode acquisition and AEC mode in PA chest X-ray. Method - 68 images (with and without lesions) were acquired using an anthropomorphic chest phantom using a Wolverson Arcoma X-ray unit. These images were compared against a reference image using the 2 alternative forced choice (2AFC) method. The effective dose (E) was calculated using PCXMC software using the exposure parameters and the DAP. The exposure index (lgM provided by Agfa systems) was recorded. Results - Exposure time decreases more when applying the 10 kVp rule with manual mode (50%–28%) when compared with automatic mode (36%–23%). Statistical differences for E between several ionization chambers' combinations for AEC mode were found (p = 0.002). E is lower when using only the right AEC ionization chamber. Considering the image quality there are no statistical differences (p = 0.348) between the different ionization chambers' combinations for AEC mode for images with no lesions. Considering lgM values, it was demonstrated that they were higher when the AEC mode was used compared to the manual mode. It was also observed that lgM values obtained with AEC mode increased as kVp value went up. The image quality scores did not demonstrate statistical significant differences (p = 0.343) for the images with lesions comparing manual with AEC mode. Conclusion - In general the E is lower when manual mode is used. By using the right AEC ionising chamber under the lung the E will be the lowest in comparison to other ionising chambers. The use of the 10 kVp rule did not affect the visibility of the lesions or image quality.por
dc.identifier.citationReis C, Gonçalves J, Klompmaker C, Bárbara AR, Bloor C, Hogg P, et al. Image quality and dose analysis for a PA chest X-ray: comparison between AEC mode acquisition and manual mode using the 10 kVp ‘rule’. Radiography. 2014;20(4):339-45.por
dc.identifier.doi10.1016/j.radi.2014.06.001
dc.identifier.urihttp://hdl.handle.net/10400.21/3739
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherElsevierpor
dc.relation.publisherversionhttp://www.sciencedirect.com/science/article/pii/S1078817414000662por
dc.subjectRadiologypor
dc.subjectImage qualitypor
dc.subjectPA chest X-raypor
dc.subjectAECpor
dc.subject10 kVp rulepor
dc.subjectEffective dosepor
dc.titleImage quality and dose analysis for a PA chest X-ray: comparison between AEC mode acquisition and manual mode using the 10 kVp ‘rule’por
dc.typejournal article
dspace.entity.typePublication
oaire.citation.endPage7por
oaire.citation.startPage1por
oaire.citation.titleRadiographypor
rcaap.rightsrestrictedAccesspor
rcaap.typearticlepor

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Image quality and dose analysis for a PA chestX-ray.pdf
Size:
673.98 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections