Paediatric phantom dose optimization using digital radiography with variation of exposure parameters and filtration whilst minimising image quality impairment

dc.contributor.authorLança, Luís
dc.contributor.authorBowdler, M. W.
dc.contributor.authorCreedon, J.
dc.contributor.authorDayer, V.
dc.contributor.authorStensholt, N.
dc.contributor.authorPinhão, S.
dc.contributor.authorVisser, M. B.
dc.contributor.authorJorge, José
dc.date.accessioned2018-12-28T20:53:22Z
dc.date.available2018-12-28T20:53:22Z
dc.date.issued2017
dc.description.abstractObjective: To induce a reduction in dose, using a paediatric phantom, through the variation of exposure parameters and filtration, without adversely affecting image quality. Methods: All images were acquired using a Kyoto Kagaku paediatric phantom and a Canon DR detector. The phantom was positioned supine for all projections: wrist (DP, lateral) and ribs (AP, oblique). Three dose protocols were established using different mAs values (high, medium and low) and copper (Cu) filtration was added to each protocol. DAP was used to calculate the ESD for each exposure. Using ImageJ, CNR was calculated for the physical measurement of image quality. Image quality was assessed by fifteen observers (visual grading analysis (VGA)). Results: The highest doses were recorded with the high dose protocol, ranging from 5.60-39.22μGy for the wrist and 5.33-129.67μGy for the ribs. When increasing the Cu filtration a decrease in ESD was observed. A difference of 0.1 in VGA score was noted between high and low dose protocols without the use of filtration, while a difference of 0.3 was noted when using filtration. As mAs increased, VGA scores increased. Fracture visibility was minimally affected by Cu filtration or projection variation. Conclusion: The variation of exposure parameters in digital radiography can achieve a dose reduction without impairing image quality in bone fractures. Superior image quality can be achieved for DP and lateral wrist projections without Cu filtration. However, the addition of Cu filtration for the rib projections has almost no impact on overall image quality.pt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.citationLança L, Bowdler MW, Creedon J, Dayer V, Stensholt N, Pinhão S, et al. Paediatric phantom dose optimization using digital radiography with variation of exposure parameters and filtration whilst minimising image quality impairment. In: Meijer A, Buissink C, Hogg P, editors. OPTIMAX 2017: optimising image quality for medical image. Oslo: Akershus University College; 2017. p. 77-92.pt_PT
dc.identifier.isbn9781912337095
dc.identifier.urihttp://hdl.handle.net/10400.21/9213
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.publisherAkershus University Collegept_PT
dc.relation.publisherversionhttp://usir.salford.ac.uk/46104/7/OPTIMAX%202017%20ed.pdfpt_PT
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/pt_PT
dc.subjectRadiologypt_PT
dc.subjectDigital radiographypt_PT
dc.subjectDosimetrypt_PT
dc.subjectPhantompt_PT
dc.subjectChildrenpt_PT
dc.subjectExposure parameterspt_PT
dc.subjectFiltrationpt_PT
dc.titlePaediatric phantom dose optimization using digital radiography with variation of exposure parameters and filtration whilst minimising image quality impairmentpt_PT
dc.typebook part
dspace.entity.typePublication
oaire.citation.endPage92pt_PT
oaire.citation.startPage77pt_PT
rcaap.rightsopenAccesspt_PT
rcaap.typebookPartpt_PT

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Paediatric phantom dose optimization using digital radiography with variation of exposure parameters and filtration whilst minimising image quality impairment.pdf
Size:
9.27 MB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: