Repository logo
 
Publication

Validação do dispositivo automático de medição da pressão arterial, OMRON® M6 Comfort, segundo o Protocolo Internacional da Sociedade Europeia de Hipertensão (2010)

dc.contributor.authorMarques, Ana
dc.contributor.authorRevige, Guida
dc.contributor.authorMarques, Inês
dc.contributor.authorSilva, Ana
dc.contributor.authorCunha, Gilda
dc.contributor.authorFonseca, Virgínia
dc.contributor.authorLobato, João
dc.date.accessioned2012-12-10T14:46:33Z
dc.date.available2012-12-10T14:46:33Z
dc.date.issued2012-11
dc.description.abstractIntrodução – A medição da pressão arterial (PA), utilizando dispositivos automáticos, é frequentemente realizada na prática clínica e na automedição, permitindo adquirir informação fiável para o diagnóstico, controlo e tratamento da hipertensão arterial. Porém, muitos dos dispositivos automáticos disponíveis no mercado não estão validados segundo protocolos existentes para o efeito. O objetivo do estudo foi confirmar a validação do dispositivo de medição automática da PA, OMRON® M6 Comfort, segundo o Protocolo Internacional da European Society of Hypertension (ESH), de 2010, para a validação de dispositivos de medição automática da PA em adultos. Metodologia – Foram estudados 33 indivíduos, aos quais foram realizadas 9 medições sequenciais da PA, no braço esquerdo, com um esfignomanómetro aneróide alternando com o dispositivo automático. Seguidamente avaliaram-se as diferenças entre os valores obtidos pelos dispositivos para a pressão arterial sistólica (PAS) e diastólica (PAD), classificando-as em três níveis (≤ 5, ≤ 10 ou ≤ 15 mmHg). O número das diferenças em cada nível foi comparado ao requerido pelo Protocolo (fase 1.1). Para cada sujeito foi ainda determinado o número de diferenças com valores ≤ 5 mmHg. Pelo menos 24 dos 33 indivíduos devem ter 2 ou 3 diferenças com valores ≤ 5 mmHg e no máximo 3 dos 33 indivíduos podem apresentar as 3 diferenças com valores > 5 mmHg (fase 1.2). Resultados – O dispositivo OMRON® M6 Comfort foi aprovado nas fases 1.1 e 1.2 para a PAS e PAD. A média das diferenças entre as medições da PA, determinada pelos dispositivos automático e manual, foi de -0,82 ± 5,62 mmHg para a PAS e 2,14 ± 5,15 mmHg para a PAD. Considerações Finais – O dispositivo OMRON® M6 Comfort é válido para a medição da PA em adultos, de acordo com o Protocolo Internacional da ESH, de 2010. - ABSTRACT - Introduction – The measurement of blood pressure (BP) using automatic devices is often performed in clinical practice and self-measurement allowing the acquisition of reliable information for the diagnosis, monitoring and treatment of hypertension. However not all of the automated devices available in the market are validated in accordance with the existing protocols for this purpose. The purpose of this study was to confirm the validation of the automatic measuring device of the BP, OMRON® M6 Comfort, according to the “European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults”. Methodology – The study involved 33 subjects, in each one of them, 9 sequential measurements of BP were performed, in the left arm, with the aneroid sphygmomanometer alternating with the automatic device. Afterwards, the differences on the values obtained by the different devices were evaluated, for systolic blood pressure (SBP) and diastolic (DBP), and these differences were then classified into three levels (≤ 5, ≤ 10 or ≤ 15 mmHg). The number of differences at each level was compared to the number required by the protocol (phase 1.1). For each subject the number of differences with values ≤ 5 mmHg was also determined. At least 24 of the 33 subjects should have 2 or 3 differences with values ≤ 5 mmHg and a maximum of 3 of the 33 subjects may have all differences with values > 5 mmHg (phase 1.2). Results – The device OMRON M6 Comfort ® was approved in phases 1.1 and 1.2 for SBP and DBP. The average difference between measurements of BP, as determined by automatic and manual devices, was -0.82 ± 5.62 mmHg for SBP and 2.14 ± 5.15 mmHg for DBP. Conclusion – The device OMRON M6 Comfort® is valid for measuring BP in adults, according to the ESH International Protocol of 2010.por
dc.identifier.citationMarques A, Revige G, Marques I, Silva A, Cunha G, Fonseca V, Lobato J. Validação do dispositivo automático de medição da pressão arterial, OMRON® M6 Comfort, segundo o Protocolo Internacional da Sociedade Europeia de Hipertensão (2010). Saúde & Tecnologia. 2012;(8):47-54.por
dc.identifier.issn1646-9704
dc.identifier.urihttp://hdl.handle.net/10400.21/1986
dc.language.isoporpor
dc.peerreviewedyespor
dc.publisherEscola Superior de Tecnologia da Saúde de Lisboapor
dc.subjectCardiologiapor
dc.subjectPressão arterialpor
dc.subjectValidaçãopor
dc.subjectDispositivo automáticopor
dc.subjectDispositivo manualpor
dc.subjectOMRON® M6 Comfortpor
dc.subjectCardiologypor
dc.subjectBlood pressurepor
dc.subjectValidationpor
dc.subjectAutomatic devicepor
dc.subjectManual devicepor
dc.titleValidação do dispositivo automático de medição da pressão arterial, OMRON® M6 Comfort, segundo o Protocolo Internacional da Sociedade Europeia de Hipertensão (2010)por
dc.title.alternativeValidation of the OMRON M6 Confort®: automatic blood pressure measuring device, according to the International Protocol of the European Society of Hypertension (2010)por
dc.typejournal article
dspace.entity.typePublication
oaire.citation.endPage54por
oaire.citation.startPage47por
oaire.citation.titleSaúde & Tecnologiapor
rcaap.rightsopenAccesspor
rcaap.typearticlepor

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Validação do dispositivo automático de medição da pressão arterial.pdf
Size:
581.17 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections