Fast-Field Cycling Nuclear Magnetic Resonance relaxometer's electromagnet with optimized homogeneity and reduced volume

Use this identifier to reference this record.
Name:Description:Size:Format: 
Fast-Field_EMargato.pdf1.14 MBAdobe PDF Download

Advisor(s)

Abstract(s)

In this article a Fast Field Cycling (FFC) Nuclear Magnetic Resonance (NMR) electromagnet with low power consumption (less than 200 W), high field homogeneity and reduced volume is projected and described. The electromagnet is iron and copper based, possessing a high permeability and allowing for good magnetic field homogeneity in the operating range of 0 to 0.33 T. With this solution, it is possible to increase 65% the maximum magnetic field keeping the magnetic field homogeneity in comparison with former similar FFC relaxometers. Electromagnet's experimental and simulation results evaluating the generated magnetic field, field homogeneity, heating effects and cooling requirements are also presented. In addition, some technical aspects of the required coupled systems such as the cooling, sample heating are assessed.

Description

Keywords

Electromagnet Relaxometer Fast Field Cycling Volume Homogeneity

Citation

VIDEIRA, P.; [et al] – Fast-Field Cycling Nuclear Magnetic Resonance relaxometer's electromagnet with optimized homogeneity and reduced volume. Renewable Energy & Power Quality Journal (RE&PQJ). ISSN 2172-038X. Vol. 17 (2019), pp. 1-7

Research Projects

Research ProjectShow more
Research ProjectShow more

Organizational Units

Journal Issue

Publisher

European Association for the Development of Renewable Energies, Environment and Power Quality (EA4EPQ)

CC License

Altmetrics