dc.description.abstractTo characterize the mean glandular dose (MGD) in a sample of healthcare providers for digital mammography in Portugal. To compare the achieved values with European references. The MGD was measured on a poly-methyl-methacrylate phantom (45 mm) for each system using dosimeters. In addition, MGD was estimated using exposure settings collected from mammography exams in a clinical context. Data were collected from 25 computed-radiography systems (CR) and 13 integrated digital (DR). For both measurements (phantom and clinical exposures), the average MGD for CR was higher compared to the DR. For CR the mean MGD was 1.85 mGy (CC projection) and 2.10 mGy (MLO projection). For DR systems the corresponding values were 1.54 mGy (CC) and 1.68 mGy (MLO). The average MGD obtained using both methods and for both technologies is within the acceptable reference range proposed by European guidelines (<2.5 mGy). Dose Reference Levels implementation should be the next step to optimise mammography practice in Portugal.pt_PT
dc.description.versioninfo:eu-repo/semantics/publishedVersionpt_PT
dc.identifier.citationReis CS, Fartaria MJ, Alves JG, Pascoal A. Portuguese study of mean glandular dose in mammography and comparison with European references. Radiat Prot Dosimetry. 2018;179(4):391-9.pt_PT
dc.identifier.doi10.1093/rpd/ncx300pt_PT
dc.identifier.urihttp://hdl.handle.net/10400.21/7922
dc.language.isoengpt_PT
dc.peerreviewedyespt_PT
dc.publisherOxford University Presspt_PT
dc.relation.publisherversionhttps://academic.oup.com/rpd/advance-article-abstract/doi/10.1093/rpd/ncx300/4802347?redirectedFrom=fulltextpt_PT
dc.rights.urihttp://creativecommons.org/licenses/by-nc-nd/4.0/pt_PT
dc.subjectMammographypt_PT
dc.subjectMean glandular dosept_PT
dc.subjectDigital mammographypt_PT
dc.subjectEuropean guidelinespt_PT
dc.subjectPortugalpt_PT
dc.titlePortuguese study of mean glandular dose in mammography and comparison with European referencespt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.endPage399pt_PT
oaire.citation.issue4pt_PT
oaire.citation.startPage391pt_PT
oaire.citation.titleRadiation Protection Dosimetrypt_PT
oaire.citation.volume179pt_PT
rcaap.rightsrestrictedAccesspt_PT
rcaap.typearticlept_PT

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Portuguese study of mean glandular dose in mammography and comparison with European references.pdf
Size:
266.43 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections