Monte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beam

dc.contributor.authorBorges, C.
dc.contributor.authorZarza-Moreno, M.
dc.contributor.authorHeath, E.
dc.contributor.authorTeixeira, Nuno
dc.contributor.authorVaz, P.
dc.date.accessioned2013-12-10T16:41:47Z
dc.date.available2013-12-10T16:41:47Z
dc.date.issued2012-01
dc.description.abstractPurpose: The most recent Varian® micro multileaf collimator(MLC), the High Definition (HD120) MLC, was modeled using the BEAMNRCMonte Carlo code. This model was incorporated into a Varian medical linear accelerator, for a 6 MV beam, in static and dynamic mode. The model was validated by comparing simulated profiles with measurements. Methods: The Varian® Trilogy® (2300C/D) accelerator model was accurately implemented using the state-of-the-art Monte Carlo simulation program BEAMNRC and validated against off-axis and depth dose profiles measured using ionization chambers, by adjusting the energy and the full width at half maximum (FWHM) of the initial electron beam. The HD120 MLC was modeled by developing a new BEAMNRC component module (CM), designated HDMLC, adapting the available DYNVMLC CM and incorporating the specific characteristics of this new micro MLC. The leaf dimensions were provided by the manufacturer. The geometry was visualized by tracing particles through the CM and recording their position when a leaf boundary is crossed. The leaf material density and abutting air gap between leaves were adjusted in order to obtain a good agreement between the simulated leakage profiles and EBT2 film measurements performed in a solid water phantom. To validate the HDMLC implementation, additional MLC static patterns were also simulated and compared to additional measurements. Furthermore, the ability to simulate dynamic MLC fields was implemented in the HDMLC CM. The simulation results of these fields were compared with EBT2 film measurements performed in a solid water phantom. Results: Overall, the discrepancies, with and without MLC, between the opened field simulations and the measurements using ionization chambers in a water phantom, for the off-axis profiles are below 2% and in depth-dose profiles are below 2% after the maximum dose depth and below 4% in the build-up region. On the conditions of these simulations, this tungsten-based MLC has a density of 18.7 g cm− 3 and an overall leakage of about 1.1 ± 0.03%. The discrepancies between the film measured and simulated closed and blocked fields are below 2% and 8%, respectively. Other measurements were performed for alternated leaf patterns and the agreement is satisfactory (to within 4%). The dynamic mode for this MLC was implemented and the discrepancies between film measurements and simulations are within 4%. Conclusions: The Varian® Trilogy® (2300 C/D) linear accelerator including the HD120 MLC was successfully modeled and simulated using the Monte CarloBEAMNRC code by developing an independent CM, the HDMLC CM, either in static and dynamic modes.por
dc.identifier.citationBorges C, Zarza-Moreno M, Heath E, Teixeira N, Vaz P. Monte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beam. Med Phys. 2012;39(1):415-23.por
dc.identifier.issn0094-2405
dc.identifier.urihttp://hdl.handle.net/10400.21/2984
dc.language.isoengpor
dc.peerreviewedyespor
dc.publisherAmerican Institute of Physicspor
dc.relation.publisherversionhttp://scitation.aip.org/content/aapm/journal/medphys/39/1/10.1118/1.3671935por
dc.subjectComputer simulationpor
dc.subjectParticle accelerators/instrumentationpor
dc.subjectRadiometry/instrumentationpor
dc.subjectRadiometry/methodspor
dc.subjectRadiotherapy dosagepor
dc.subjectDosimetrypor
dc.subjectRadiotherapy planning, Computer-Assisted/methodspor
dc.subjectRadiotherapy, High-Energy/instrumentationpor
dc.titleMonte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beampor
dc.typejournal article
dspace.entity.typePublication
oaire.citation.endPage423por
oaire.citation.startPage415por
oaire.citation.titleMedical Physicspor
oaire.citation.volume39por
rcaap.rightsrestrictedAccesspor
rcaap.typearticlepor

Files

Original bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
Monte Carlo modeling and simulations of the High Definition.pdf
Size:
301.91 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections