Utilize este identificador para referenciar este registo: http://hdl.handle.net/10400.21/2985
Título: Dosimetric effect by shallow air cavities in high energy electron beams
Autor: Zarza-Moreno, M.
Carreira, P.
Madureira, L.
Miras del Rio, H.
Salguero, F. J.
Leal, A.
Teixeira, Nuno
Jesus, A. P.
Mora, G.
Palavras-chave: Physics
Air cavity
EBT2 Gafchromic film
Electron beam
Monte Carlo algorithms
Pencil beam algorithms
Data: Mar-2014
Editora: Elsevier
Citação: Zarza-Moreno M, Carreira P, Madureira L, Miras del Rio H, Salguero FJ, Leal A, Teixeira N, el al. Dosimetric effect by shallow air cavities in high energy electron beams. Phys Med. 2014;30(2):234-41.
Resumo: This study evaluates the dosimetric impact caused by an air cavity located at 2 mm depth from the top surface in a PMMA phantom irradiated by electron beams produced by a Siemens Primus linear accelerator. A systematic evaluation of the effect related to the cavity area and thickness as well as to the electron beam energy was performed by using Monte Carlo simulations (EGSnrc code), Pencil Beam algorithm and Gafchromic EBT2 films. A home-PMMA phantom with the same geometry as the simulated one was specifically constructed for the measurements. Our results indicate that the presence of the cavity causes an increase (up to 70%) of the dose maximum value as well as a shift forward of the position of the depthedose curve, compared to the homogeneous one. Pronounced dose discontinuities in the regions close to the lateral cavity edges are observed. The shape and magnitude of these discontinuities change with the dimension of the cavity. It is also found that the cavity effect is more pronounced (6%) for the 12 MeV electron beam and the presence of cavities with large thickness and small area introduces more significant variations (up to 70%) on the depthedose curves. Overall, the Gafchromic EBT2 film measurements were found in agreement within 3% with Monte Carlo calculations and predict well the fine details of the dosimetric change near the cavity interface. The Pencil Beam calculations underestimate the dose up to 40% compared to Monte Carlo simulations; in particular for the largest cavity thickness (2.8 cm).
Peer review: yes
URI: http://hdl.handle.net/10400.21/2985
ISSN: 1120-1797
Versão do Editor: http://www.sciencedirect.com/science/article/pii/S1120179713002366
Aparece nas colecções:ESTeSL - Artigos

Ficheiros deste registo:
Ficheiro Descrição TamanhoFormato 
Dosimetric effect by shallow air cavities in high energy electron beams.pdf1,85 MBAdobe PDFVer/Abrir


FacebookTwitterDeliciousLinkedInDiggGoogle BookmarksMySpace
Formato BibTex MendeleyEndnote Degois 

Todos os registos no repositório estão protegidos por leis de copyright, com todos os direitos reservados.